TLD-3500 Automatic TLD Reader

Contact Us
  • Fast And Safe Logistics
  • Get samples for free

The TLD-3500 Automatic TLD Reader is used for periodic monitoring of personal external radiation dose. The equipment automatically measures the thermoluminescent dose meter worn by personnel, accurately and reliably evaluates personal dose data, and detects and collects data by connecting to computer application software and database to form a complete dose file and report.

Physical parameters
Dimensions, weight450*400*400mm; 30 kg;
Shell materialSheet metal shell, spray or anodized
Environment-20℃--50℃, 10%-60% humidity
Power supply:  220VAC/50Hz
Operating parameters
Operation modeDosimeter is installed as a whole, automatic and continuous measurement
Detection modePhotomultiplier tube detection, analog signal amplification processing
Interaction modeWith touch screen human-machine interface and application software synchronous dual operation
Maintenance modeWith self-diagnosis function; modular design facilitates rapid replacement on site;
Calibration modeIrradiator can be built in, periodic calibration of equipment and dosimeter screening
Adapting dosimeterWhole body dosimeter (four dose elements)
Measurement of radiationγ, β, X, neutron rays
Heating unit parameters
Heating methodElectrode heating, linear heating, maximum temperature 500℃
Acquisition temperatureRoom temperature ~ 500℃ can be set arbitrarily
Preheating temperatureRoom temperature ~ 200℃ can be set arbitrarily
Heating rate1℃/s ~ 50℃/s can be set arbitrarily
Temperature control accuracy±2℃
Test unit parameters
Measurement methodPhotomultiplier tube (PMT), analog signal amplification 109 order of magnitude
High voltage supplyHigh voltage 0~1500V adjustable, high voltage stability ≤1%
Detection unit coolingSemiconductor plus controlled cooling, cooling temperature: 8~20℃ freely settable
Measurement range0.1μSv~10Sv; up to 300Sv (sensitized magnesium titanium sheet)
Detection lower limit0.1μSv
Measurement resolution0.1uSv after full preheating under laboratory conditions
Measurement stability10 consecutive measurements after full preheating, standard deviation RSD≤1%;
Dark currentContinuous multiple readings after preheating RSD≤1%
Reference light Continuous multiple readings after preheating RSD≤1%