ICP-MS 2000 Inductively Coupled Plasma Mass Spectrometry
Introduction:
Inductively Coupled Plasma Mass Spectrometry(ICP-MS),has a microscale(10-6), trace(10-9)、ultra trace(10-12) element analyses technique. It can analyze most of the elements in the periodic table. It has very low detection limit, wide trends linearity range, lew interference, high accuracy, high analyzed speed, and it also can analyze isotope.
ICP-MS2000 is the first inductively coupled plasma mass spectrometer in China. lt can satisfy the clientele requirements and has high performance-price ratio. It mainly applies to environmental、foodstuff、 semiconductor, medicine and physiological analysis、 nuke industry etc.
The network interface for serial inductance, enhanced resistance to interference, improve the data transmission efficienc0and stability
Very kinds of protect function
The software is easy operated. No special training
The ion lens, Quadrupole are easily installing, dismantling and maintain
Perfectly technique of electromagnetic shielding can reduce electromagnetic interference (EM)0Consuming material of instrument is developed by shuoboda itself, it is cost-effective
Hich grade sale service10minutes respond,48hours door-to-door service, customer service center follow-up service at all times、ensure service quality
Product performance superiority:
Quick analyze speed, simple operate, high sensitivity, low background noise, good effect of eliminating interference, easy maintain
OneClick plasma setting made plasma optimizing more convenient and excellent reproducibility
Advanced plasma shielding Technology greatly improve the instrument response rate and limit of low mass number, achieve ppt level
lt has particular Activity connector structure, and it can replace, install, dismantling sampling cone and skimmer cone in empty space, it's convenient for daily maintenance
The ion efficiency of transmission and eliminate interference capability of polyatomic ion is improved by using the hexapole collision cell
It doesn’t need Digital/Analogue Conversion, controlled by computer and can achieve consistence dynamic range to 9 orders of magnitude
The new type Vacuum Chambers structure has no wire to connect. Each module uses dissymmetric(al) and plug-in installing
Main apply field:
Environment field: drinking water、seawater、 environment water resource foodstuff、hygiene and disease control, commodity inspection etc.
Semiconductor field: high pure metal, high purity reagent, ultra trace impurity of Si wafe, photoresist etc.
Medicine and physiological analysis field: Medical Research of hair、 whole blood、 serum、 urine sample, biological organization etc. Especially the test of Pb in whole blood
Nuke industry field: analysis of nuclear fuel radioisotope, pollution of Primary cooling water etc.
Other field: as chemical industry, lithification, geology etc.
Technical parameters:
Mass range:2~255 amu
Linear range: ≥108
Sensitivity: Be≥2x106, In≥35x106, U≥30x106 unit (cps/mg/L)
Detection limit: Be≤10; In≤ 2;U≤2 unit(ng/L)
Resolution:0.6~0.8 amu
SNR(Signal to Noise Ratio): ≥50x106
Background noise: ≤2 cps (all mass number range)
Mass axis stability: ≤0.05 amu/24 h
Stability RSD: short period≤3%; long time≤4%
Oxide ion: CeO+/Ce+≤3%
Divalent ions.69Ba2+/138Ba ≤3%
lsotope ratio: (107Ag/109Ag) ≤0.3%
Abundance sensitivity: ≤1x10-6 low mass number; ≤5x10-7 high mass number
Software advantage:
ICP-MS2000 provides the most convenient operating software, very intuitive, comprehensive. The software contains all current analysis methods, including a special isotope ratio and isotope dilution method.
Intelligent selection method, intelligent instrument tuning, QC, various analysis methods combination function, sequence analysis, the function of automatic monitoring, custom report format.