BD-D92-3536W fully automatic open flash point and fire point tester is suitable for testing in accordance with the national standard GB/T3536 "Determination of Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Method)" and the national industry standard JTJ052 "Testing Procedures for Highway Engineering Asphalt and Asphalt Mixtures" The method specified in T0611 "Bitumen Flash Point and Fire Point Test (Cleveland Open Cup Method)"

It is used to determine the flash point and fire point of petroleum products and asphalt with the exception of fuel oil and open cup flash points lower than 79°C.

Asphalt ignition point is a reference index for construction safety, so the instrument includes both methods and indicators at the same time. This instrument can be widely used in the petroleum, railway, aviation, electric power industry, highway industry; and related colleges and universities, scientific research institutes, oil testing, asphalt testing and other units and departments, for testing the open cup flash point and Randian is an ideal substitute for similar imported instruments.

1. This instrument adopts a large color LCD screen display, a touch screen keyboard, and a full Chinese human-machine dialogue interface. It has prompt menus and guided input methods for parameters such as expected flash point, sample label, atmospheric pressure, and test date.

2. Simulation tracking displays the function curve of temperature rise and test time. It has a Chinese incorrect operation software prompt modification function, and is equipped with a parameter prompt function such as test date and test time.

3. It is a dual-purpose machine that can automatically detect the flash point and ignition point of the sample as required. During the process of detecting the flash point of asphalt, if a film is formed on the liquid surface, the instrument has the function of automatically removing the film to detect the flash point.

4. Equipped with RS232 serial communication interface, which can be connected to USB and store 256 sets of historical data; it can communicate with the computer and save test results in large capacity; it can also realize network remote communication function for special orders.

5. Automatically correct the influence of atmospheric pressure on the test and calculate the correction value.

6. Differential detection and automatic correction of system deviations.

7. Scanning, detecting, and printing data are automatically completed, and the test arm automatically rises and falls.

8. Electronic ignition gas flame, automatic forced air cooling at the end of each test, shortening the waiting time.

9. Designed with a safety fire extinguishing cover, the fire extinguishing cover automatically completes the fire extinguishing function after the ignition point detection is completed.

1. Temperature measurement: Range: room temperature ~ 400℃, temperature display unit: 0.1℃

2. Repeatability: ≤8℃ (flash point and fire point)

3. Reproducibility: ≤17℃ (flash point), ≤14℃ (fire point)

4. Heating speed: in line with GB/T3536 standard

5. Ignition method: gas flame, ignited by electronic ignition

6. Communication interface: With RS232 serial communication interface

7. Usage environment: temperature (10~40)℃, relative humidity ≤80%

8. Working power supply: AC (220±10%) V, 50Hz

9. The power consumption of the whole machine: no more than 500W

10. Instrument dimensions: 460 ㎜×430 ㎜×290 ㎜ (length×width×height)

11. Net weight of the instrument: 22.5kg