The DA-2000E thermoluminescence dose TLD reader can test the thermoluminescence detector irradiated by X, γ, β, neutron and other rays.
It provides the reading value and dose value, with automatic calibration and deduction Background, heating parameter setting and printing functions.
The thermoluminescent dose reader can be connected to a computer for data management, printing, etc. through the thermoluminescent dose reader application software.
It is suitable for environmental background measurement to radiation therapy dose, radiation flaw detection exposure dose control, nuclear accident dose measurement and monitoring, etc.
Light measurement system | |
Dose measurement range | LiF:Mg,Ti:10-5Gy ~ 1Gy LiF:Mg,Cu,P: 10-7Gy ~ 10Gy LiF:Mg,Ti-M: 0~100Gy |
Measurement items | X, γ, β, neutron |
system stability | ≤0.2% |
Reading time | 40s(can be customized) |
Variation factorization of sensitivity repeatability | ≤0.1%±0.05%/℃ |
Drawer light leakage | ≤0.1% of the internal light source |
background value | ≤300 |
high pressure | 0~1500V |
power supply | AC 220 V±10%, 50Hz |
Heating system | |
Heating temperature range | RT~400℃ |
Heating temperature repeatability | ≤1% |
Heating temperature deviation | ≤±2℃ |
heating time | 0~500s |
Heating time repeatability | ≤0.1% |
Heating rate | 1~40℃·s-1 |
Conditions of Use | |
working environment | indoors |
Working temperature | 10~40℃ |