Professional Supplier of Laboratory Scientific Instruments

Environmental Monitoring

AMS02 Online Aerosol Intelligent Monitoring System (Unattended)

Contact Us
  • Fast And Safe Logistics
  • Get samples for free

Online Aerosol Intelligent Monitoring System

Product introduction           

Combination of α, β and γ measurements in the system, 4 detectors used in the aerosol system, 3 in-line measurement positions and a separate additional measurement position are possible with special filter material for iodine evaluation (at increasing concentrations In case of activation), to achieve efficient and reliable monitoring of artificial nuclides in the air.

• Multiple detectors are optional, high-purity germanium, lanthanum bromide, sodium iodide, plastic scintillation detector, RS04, GSP02

• The station integrates 5 detectors working together

• There are three kinds of filter devices, two kinds of filter membranes and a filter tank with lead shielding to filter and measure aerosol, inorganic iodine and organic iodine respectively.

• Online monitoring, all monitoring data and alarm information are sent in real time.

• Fully automatic aerosol monitoring, regular self-inspection to ensure reliable measurement data

• High intelligence and stability of the machine gun can be unattended around the clock, and its consumables can be used continuously for one year

• Optional weather system matching use

• Cooperating with SCADA server, multiple units can be networked and operated visually. The comprehensive radioactivity monitoring data of each important location can be seen at a glance, and the alarm information and operation status of each monitoring data can be grasped in real time.

Technical Parameters

• Dimensions: 730 x 920 x 1520 (2100) mm

• Weight: Approx. 415 kg

• Power supply: 230 V AC / 50 Hz / 950 VA

• Environment: Temperature -15°C ~ + 25°C

• Collection gas: temperature: -30°C ~ + 40°C

• Relative humidity: 0~99%

Main detector

• -2” x 2” Na(Tl) (2 pcs)

Resolution 8% (137Cs 662 keV)

• -PIPS 1700 mm2

Resolution 55 keV (α241Am) <30 keV (β)

• - Charge-cooled coaxial germanium detector (HPGe) - no LN2 required Resolution 2.0 keV FWHM at 1.33 MeV

Relative efficiency 33% at 1.33 MeV

• -RS04 dose rate detector

Measuring range: 10nSv/h~10Sv/h Temperature: -30°C~+70°C

Measurement uncertainty: [H*(10)] ≤ 1 Sv/h: ±10%

[H*(10)] > 1 Sv/h: ±15%

Get in Touch With Us

Experienced service team and strong production support team provide client worry-free order service.

    Need help? Contact us.

    If you are interested in any of our products or would like to discuss a customized order, Please feel free to contact us.

    Contact Us