The TLD-3500 Automatic TLD Reader is used for periodic monitoring of personal external radiation dose. The equipment automatically measures the thermoluminescent dose meter worn by personnel, accurately and reliably evaluates personal dose data, and detects and collects data by connecting to computer application software and database to form a complete dose file and report.
Physical parameters | |
Dimensions, weight | 450*400*400mm; 30 kg; |
Shell material | Sheet metal shell, spray or anodized |
Environment | -20℃--50℃, 10%-60% humidity |
Power supply: | 220VAC/50Hz |
Operating parameters | |
Operation mode | Dosimeter is installed as a whole, automatic and continuous measurement |
Detection mode | Photomultiplier tube detection, analog signal amplification processing |
Interaction mode | With touch screen human-machine interface and application software synchronous dual operation |
Maintenance mode | With self-diagnosis function; modular design facilitates rapid replacement on site; |
Calibration mode | Irradiator can be built in, periodic calibration of equipment and dosimeter screening |
Adapting dosimeter | Whole body dosimeter (four dose elements) |
Measurement of radiation | γ, β, X, neutron rays |
Heating unit parameters | |
Heating method | Electrode heating, linear heating, maximum temperature 500℃ |
Acquisition temperature | Room temperature ~ 500℃ can be set arbitrarily |
Preheating temperature | Room temperature ~ 200℃ can be set arbitrarily |
Heating rate | 1℃/s ~ 50℃/s can be set arbitrarily |
Temperature control accuracy | ±2℃ |
Test unit parameters | |
Measurement method | Photomultiplier tube (PMT), analog signal amplification 109 order of magnitude |
High voltage supply | High voltage 0~1500V adjustable, high voltage stability ≤1% |
Detection unit cooling | Semiconductor plus controlled cooling, cooling temperature: 8~20℃ freely settable |
Measurement range | 0.1μSv~10Sv; up to 300Sv (sensitized magnesium titanium sheet) |
Detection lower limit | 0.1μSv |
Measurement resolution | 0.1uSv after full preheating under laboratory conditions |
Measurement stability | 10 consecutive measurements after full preheating, standard deviation RSD≤1%; |
Dark current | Continuous multiple readings after preheating RSD≤1% |
Reference light | Continuous multiple readings after preheating RSD≤1% |