Professional Supplier of Laboratory Scientific Instruments
Professional Supplier of Laboratory Scientific Instruments
SPD-H5O Lab Inert Loop Spray Dryer is specially for organic solvents, the inert gas medium formed by circulating nitrogen is used to replace the usual air medium for spray drying. In addition to recovering products, organic solvents can also be recovered.
This closed-loop spray system is specially designed for the low flash point or slightly toxic, flammable and explosive characteristics of organic solvents. It is an ideal closed-loop dryer for laboratory small-scale to pilot-scale tests, and provides a new and reliable method for drying laboratory organic solvents, toxic materials and easily oxidizable materials.
Capacity | 5000ml/h | Range | 30-250℃ | Gas | N2 |
No. | Item Name | Specifications |
1. | Inlet air temperature | 30~250℃(accuracy ±0.5℃) |
2. | Minimum feed amount | 50ml |
3. | Drying time | 0.8s~1.0s |
4. | Feed speed | 50~5000ml/h |
5. | Atomization structure | Electric centrifugal atomizer |
6. | Oxygen measurement system detection range | 0-100%(accuracy 0.1) |
7. | Oxygen detection module | Ensure that the oxygen concentration is below 3% before feeding (value can be set) |
8. | Pressure measurement system range | 0-100Kpa |
9. | Explosion-proof module | Automatic pressure relief above 3Kpa (value can be set) |
11. | Explosion-proof membrane | Automatic type |
12. | Air volume adjustment | 0-320 m3/h arbitrary adjustment (variable frequency fan) |
13. | Heating power | 10kw |
14. | Overall machine size | 1600*1400*2100mm |
15. | Power supply | 380V |
Experienced service team and strong production support team provide client worry-free order service.
If you are interested in any of our products or would like to discuss a customized order, Please feel free to contact us.
Contact Us