BD-D525-8018D-1 fully automatic gasoline oxidation stability tester is designed and manufactured according to the requirements stipulated in the People's Republic of China standard GB/T 8018-2015 "Gasoline Oxidation Stability Determination Method (Induction Period Method)" and is suitable for In accordance with the requirements specified in the GB/T 8018-2015 standard, the oxidation stability of gasoline under accelerated oxidation conditions was determined.
Meets testing requirements specified in ASTM D525 standard.
1. Desktop structure, integrated design of the whole machine, the test part and the control part are combined into one, high integration, no need for external computer, mouse and keyboard.
2. It adopts built-in industrial computer, works in fully automatic mode, 10.1-inch color touch screen, Windows7 operating system, full Chinese interface, simple and clear, beautiful appearance and easy to operate.
3. Change the traditional water bath to a metal bath, which is pollution-free, does not require water replenishment, and is more convenient to operate.
4. Double-hole design can test two samples at the same time, making it more convenient for users to conduct parallel sample tests.
5. The complete thermal insulation system not only saves energy but also avoids the risk of burns to the operator.
6. The oxygen bomb and test system are integrated in design, and equipped with an automatic pressure relief protection device for bomb body over-pressure, making it safer to use.
7. The color display screen displays two oxygen bomb test curves, time and pressure correspondence data, two oxygen bomb synthesis comparison curves, oxygen bomb pressure and temperature real-time data, etc. in real time. The timeline of the curve will automatically change accordingly. It varies depending on the actual curve time length, up to 3000 minutes. It has a wide adaptability range and the test conditions are intuitive and clear.
8. The software design has a high degree of automation and can automatically complete a series of operations. The human-machine dialogue interface is friendly and guides users to correctly execute the operation process and avoid errors.
9. The design of the oxygen bomb positioning holes on both sides makes it convenient for the operator to disassemble the oxygen bomb cover in the positioning hole, and also facilitates the placement of the oxygen bomb before and after the test.
1. Metal bath temperature control point: 100.0℃
2. Metal bath temperature control accuracy: ±0.1℃
3. Digital precision pressure gauge range: 0~1600Kpa
4. Digital precision pressure sensor accuracy: ±2‰
5. LCD touch screen: 10.1
6. Operation method: WINDOWS operating system, with wireless WIFI function
7. Working power supply: AC (220±10%) V, (50 ±1) Hz
8. Heat pipe power: 1400W
9. Overall dimensions: 630mm×280mm×380mm
10. Net weight of the instrument: 50kg