Professional Supplier of Laboratory Scientific Instruments

TOC Analyzer

TOC-1.0 Portable TOC (total organic carbon) Analyzer (Offline Test & cleaning validation)

Contact Us
  • Fast And Safe Logistics
  • Get samples for free

TOC-1.0 TOC tester is designed for the online and offline test of pharmaceutical grade water, WFI (water for inj.), ultra-pure water, and deionized water and etc. Based on the conductivity difference technology, it is being of high-precision capacity and short response-time, also meets the regulation standards issued by government.

Meet the requirements of TOC testing regulated by the Chinese Pharmacopoeia (2010 version) for pharmaceutical grade water

Pharmaceutical water (purified water, water for injection) on-line monitoring and laboratory testing and cleaning validation;

Environmental testing, electronic industry, food industry and so on.

This tester adapts the UV-oxidation method, to oxide the sample organic contaminants into carbondioxide gasthoroughly which dissolve into liquid phase instantly,this will result the conductivity difference signal between test tube and control tube,and the final TOCvalue can be converted out by the below formula:TOC=TC(total carbon in test tube)- TIC(total inorganic carbon incontrol tube).

  1. Low costs of operating ,the whole process does not require the use of any supplies, such as working gas, reagents, and other parts.
  2. Friendly interface ,easy and convenient operation
  3. Two mode, optional online or offline mode with the choice of automatic injector.
  4. Large capacity memory disk.
  5. Portable design, suitable for use in variety occasions.
  6. Automatic alarm, under the circumstance of sample TOC concentration exceeding limit.
  7. Electronicsignature,aswellasusergrademanagement.
  8. Good reproducibity of testing, supported by the high-precision sampling chip.
  9. GMP & FDA certificated, meets well with the regulation of Chinese Pharmacopoeia(2010version) issued by government.

1.Delay line

2.  UVlamp and quartz tube

3.Conductivity sensor

4.Conductivity sensor

5.Pump

ModelTOC-1.0
Power requirements / power100-240VAC, 50HZ, 120W
TOC detection range1μg / L ~ 1000 μg / L (1~1000ppb)
Conductivity detection range0.055μs / cm-8.000μs / cm
Detection limit0.001mg / L
Accuracy≤3% test range
Repeatability error≤ 3%
Zero drift≤ ± 5%
Sample temperature1- 99 ℃
Ambient temperature5-65 ℃
Power / Frequency100-240VAC / 50HZ; 120W
Dimensions410 * 240 * 310mm
equipment weight8.5kg
DisplayColor touch screen
Resolution0.001mg / L
Measuring range0.001mg / L-1mg / L (1000ppb)
measurement accuracy± 3%
Response timeWithin 5 minutes
Scope of applicationOffline laboratory, cleaning verification
Audit trail≥5 years storage
authority managementUsername & password login, level 4 permissions, meeting FDA 21 CFR PART 11
Print functionExternal mini printer
history record≥5 years storage
data backupSupport U disk export data
DisplayColor touch screen
Overall dimensions (mm)400 × 240 × 270 (length × width × height)
PC softwareable to provide
ModelTOC-1.0TOC-2.0TOC-3.0
Temperature measurement
Online analysis
Offline analysis
Zero calibration
Adaptive verification
Pipe flushing
Pump tube replacement  
Automatic decompression 
Control output  
Telecommunication   
Wireless alarm   
Audit trail 
Auto sampler 

● Standard ◎Optional

Get in Touch With Us

Experienced service team and strong production support team provide client worry-free order service.

    Need help? Contact us.

    If you are interested in any of our products or would like to discuss a customized order, Please feel free to contact us.

    Contact Us